Skip to content
Exams
Emergency
Intensive Care
Anaesthesia
Surgery
Internal Medicine
General Practice
Other Specialties
Study Guides
Practice and Tools
Start free trial
Home  /  ANZCA Primary  /  Study notes  /  Measurement methods in anaesthesia

Measurement methods in anaesthesia

ANZCA Primary LO BT_SQ 1.6 2,115 words
Free preview. This study note covers learning objective BT_SQ 1.6 from the ANZCA Primary curriculum. Inside Primex you get AI-graded SAQ practice on this topic, voice viva with the AI examiner, MCQs across the full syllabus, and a curriculum tracker that ticks off every learning objective.

SI Units and Pressure Measurement

SI Units in Anaesthesia

The International System of Units (SI) provides a standardised framework for measurement. Key SI units relevant to anaesthesia include:

Quantity SI Unit Symbol Common Clinical Unit
Force Newton N ,
Pressure Pascal Pa mmHg, cmH₂O, kPa
Temperature Kelvin K °C
Length Metre m cm, mm
Mass Kilogram kg g
Time Second s min

Pressure is defined as force per unit area:

$$P = \frac{F}{A}$$

The SI unit of pressure is the pascal: $1\ \text{Pa} = 1\ \text{N·m}^{-2}$

Because 1 Pa is a very small pressure, the kilopascal (kPa) is more clinically practical. Multiple pressure units are used in medicine, requiring fluency in conversion:

Unit Symbol Equivalent in kPa
Kilopascal kPa 1
Millimetres of mercury mmHg 0.133
Centimetres of water cmH₂O 0.0981
Atmosphere atm 101.325
Bar bar 100
Torr torr 0.133
Pounds per square inch psi 6.89

Blood pressure is conventionally expressed in mmHg; airway pressures in cmH₂O; gas cylinder pressures in kPa or bar.


Measurement of Volumes, Flows, and Pressures

Pressure Gauges and Transducers

Pressure transducers convert a mechanical pressure signal into an electrical signal. In invasive arterial blood pressure monitoring, a fluid-filled column transmits pressure to a strain-gauge transducer. Key components and sources of error are discussed under invasive blood pressure measurement.

Flowmeters

Variable-orifice (Rotameter) flowmeters use a bobbin suspended in a tapered tube. Flow is read at the top of the bobbin. These are gas-specific due to dependence on gas viscosity (at low flows, laminar, viscosity-dependent) and density (at high flows, turbulent, density-dependent).

Pneumotachograph

The pneumotachograph is a fixed-orifice, variable-pressure flowmeter. It measures the pressure drop across a known resistance (a mesh or bundle of parallel tubes):

$$\dot{V} = \frac{\Delta P}{R}$$

Wright Respirometer


Measurement of Blood Pressure

Non-Invasive Blood Pressure (NIBP)

Auscultatory method (sphygmomanometry):

Oscillotonometry (DINAMAP):

Mean arterial pressure (MAP):

$$\text{MAP} = \text{DBP} + \frac{1}{3}(\text{SBP} - \text{DBP})$$

Or equivalently: MAP = diastolic pressure + one-third of pulse pressure.

Finapres system: Continuous non-invasive measurement using a finger cuff with infrared plethysmography; maintains constant finger volume via servo-controlled cuff pressure.

Invasive Arterial Blood Pressure (IABP)

Common sites: Radial (most common), brachial, femoral arteries. Allen's test performed before radial cannulation.

System components: Arterial cannula → fluid-filled tubing → pressure transducer → amplifier/display

Zeroing: Transducer levelled at the phlebostatic axis (mid-axillary line, 4th intercostal space); opened to atmosphere to zero.

Sources of error:

Error Source Effect
Air bubbles in tubing Overdamping → underestimates systolic, overestimates diastolic
Excessive tubing length Overdamping
Catheter whip/resonance Underdamping → overestimates systolic
Incorrect zeroing/levelling Systematic offset

Fourier analysis: The arterial waveform is a complex periodic wave; accurate reproduction requires the system's natural frequency to be well above the fundamental frequency of the pulse (~1-3 Hz), with adequate damping coefficient (~0.64 optimal).

Additional waveform information: Stroke volume estimation, fluid responsiveness assessment, systolic pressure variation.

Complications: Haematoma, arterial thrombosis, distal ischaemia, infection, accidental intra-arterial drug injection.


Measurement of Cardiac Output

Methods include:


Temperature Measurement

Method Principle Clinical Use
Thermistor Resistance decreases with temperature PA catheter, core temp probes
Thermocouple Two dissimilar metals generate EMF proportional to temperature difference Skin, oesophageal
Platinum resistance thermometer Resistance increases with temperature (precise) Laboratory standard
Bimetallic strip Differential expansion of two metals Simple dial thermometers
Infrared tympanic Detects infrared emission from tympanic membrane Clinical spot measurement
Liquid crystal Temperature-dependent colour change Skin (forehead strips)

Core temperature sites (in order of accuracy): Pulmonary artery > oesophagus > nasopharynx > rectum > tympanic membrane > bladder.


Electrocardiography (ECG)

Sources of interference:

Clinical utility: Rhythm monitoring, ST-segment analysis (ischaemia), electrolyte disturbance detection, pacemaker function.


Oximetry (Pulse Oximetry)

Principles

Pulse oximetry exploits the Beer-Lambert law: light absorption is proportional to the concentration of absorbing species and path length.

Two wavelengths of light are used:

The ratio of absorbances at these two wavelengths (R/IR ratio) is used to calculate $SpO_2$. The pulsatile component of absorbance is isolated to distinguish arterial blood from venous blood and tissue.

Accuracy

Clinically accurate within the range 70-100% $SpO_2$.

Sources of Error

Cause Effect on SpO₂
Carbon monoxide (carboxyhaemoglobin) False over-reading (reads ~100%)
Methaemoglobinaemia Reads towards 85% regardless of true saturation
Nail polish (especially dark colours) Under-reading
Motion artefact Inaccurate readings
Poor perfusion/vasoconstriction Unreliable signal
Anaemia alone (without hypoxia) Minimal effect on accuracy
Bright ambient light False readings

Key limitation: $SpO_2$ does not reflect $PaCO_2$ or ventilation adequacy.

Terminology:


Infrared Gas Analysis and Capnography

Principles of Infrared Absorption

Molecules with two or more different atoms absorb infrared radiation at characteristic wavelengths (due to molecular vibration/rotation). Homodiatomic gases (O₂, N₂) cannot be measured by infrared analysis.

Key absorption wavelengths:

Gas Wavelength
CO₂ 4.26 µm (also 4.3 µm quoted)
Nitrous oxide 4.6 µm
Volatile anaesthetic agents 3.3 µm

The Beer-Lambert law applies: absorbance is proportional to gas concentration and path length. Glass absorbs infrared radiation and cannot be used for windows in analysers (sapphire or calcium fluoride used instead). Nitrous oxide can interfere with CO₂ analysis due to overlapping absorption spectra in some designs.

Capnography

Mainstream analysers: Sensor placed directly in the airway circuit; no sampling delay, but adds dead space and weight to circuit; cannot simultaneously measure other gases easily.

Sidestream analysers: Gas sampled from circuit via fine tubing at 50-200 mL/min; delay of up to 1-2 seconds acceptable (less than 38 seconds); lighter, can measure multiple gases simultaneously; prone to sampling tube occlusion/leaks and water vapour interference.

Normal Capnograph Waveform

In healthy lungs: $EtCO_2 \approx PaCO_2$ (EtCO₂ typically 2-5 mmHg less than PaCO₂). In COPD, the alveolar plateau slopes upward due to uneven emptying.

Clinical uses of capnography:

Photoacoustic spectroscopy: An alternative infrared detection method using a microphone to detect pressure waves generated by absorbed infrared energy; more stable than conventional thermal detection.


Paramagnetic and Fuel Cell Oxygen Analysis

Paramagnetic Oxygen Analyser

Oxygen is weakly paramagnetic (attracted into a magnetic field) due to two unpaired electrons. Most other anaesthetic gases are diamagnetic (weakly repelled).

Differential pressure paramagnetic analyser (modern standard):

  1. Two chambers separated by a sensitive pressure transducer
  2. Sample gas delivered to measuring chamber; reference gas (room air, 21% O₂) to reference chamber
  3. Electromagnet switched on/off at ~100-110 Hz, creating an alternating magnetic field
  4. Oxygen molecules attracted into the magnetic field create a pressure difference proportional to oxygen concentration
  5. Differential pressure measured and displayed as % O₂

Fuel Cell (Galvanic Cell)

Components: Lead anode, gold cathode, potassium hydroxide electrolyte, oxygen-permeable membrane (Teflon).

Reactions:

Current generated is proportional to oxygen partial pressure. Calibrated using 100% O₂ and room air (21% O₂).

Feature Fuel Cell Paramagnetic
Response time 20-30 s (slow, membrane diffusion) 10-20 ms (fast)
Lifespan ~1 year ~3 years
Breath-to-breath No Yes
Water vapour effect Minimal Minimal
Position Common gas outlet Common gas outlet

Basic Pulmonary Function Tests

Spirometry

Spirometry measures lung volumes and flow rates using a pneumotachograph or rolling-seal spirometer.

Measurement Definition Normal Value
FVC Forced vital capacity ~4.0-5.0 L (male)
FEV₁ Forced expiratory volume in 1 second ~3.0-4.0 L (male)
FEV₁/FVC ratio Tiffeneau index >0.7 (70%)
PEFR Peak expiratory flow rate 400-600 L/min

Obstructive pattern: FEV₁/FVC < 0.7, FVC relatively preserved (COPD, asthma) Restrictive pattern: FVC reduced, FEV₁/FVC ≥ 0.7, TLC reduced

Flow-Volume Loops

Graphical display of flow vs. volume during forced inspiration and expiration. Characteristic patterns for obstruction (scooping of expiratory limb), restriction (narrow but normal shaped), and fixed upper airway obstruction (plateau in both phases).

Wright Respirometer

Used clinically to measure tidal volume and minute ventilation at the bedside or intraoperatively. Best positioned on the expiratory limb. Under-reads at low flows; over-reads at high flows.


Clinical Relevance

Measurement Key Anaesthetic Implication
NIBP Cuff size selection critical; oscillotonometry can fail in arrhythmias
IABP Allows beat-to-beat monitoring, waveform analysis; requires zeroing at phlebostatic axis
Pulse oximetry Does not detect hypoventilation; CO poisoning causes false reassurance
Capnography Mandatory for confirming intubation; best disconnection alarm; EtCO₂ underestimates PaCO₂ in V/Q mismatch
Paramagnetic O₂ Standard for FiO₂ monitoring; fast enough for breath-to-breath analysis
Fuel cell Slower response; limited lifespan (~1 year); adequate for inspired O₂ monitoring
Infrared analysis Cannot measure O₂; nitrous oxide interference must be accounted for
Temperature Core temperature monitoring essential for prolonged anaesthesia and cardiopulmonary bypass
ECG Monitoring mode preferred intraoperatively; diagnostic mode for ischaemia detection
Spirometry Pre-operative risk stratification; FEV₁/FVC distinguishes obstructive from restrictive disease

Pressure transducer zeroing errors are a common and important source of error in invasive monitoring: for every 10 cm of incorrect height, approximately 7.4 mmHg error is introduced. All invasive pressure measurements must be zeroed at the phlebostatic axis before clinical decision-making.


Sources

Primex

Practice this topic in the app

Sit a graded SAQ on this exact LO, run a voice viva with the AI examiner, or work through MCQs that map to BT_SQ 1.6. Your free trial covers all 21 exams.

Start 7-day free trial

Quick recall flashcards

A small sample of the deck for this topic. Tap a question to reveal the answer. The full deck and spaced-repetition scheduler live inside Primex.

What is the SI unit of pressure, and how is it defined?

The pascal (Pa), defined as one newton per square metre (1 Pa = 1 N·m⁻²). In clinical practice, kilopascals (kPa) are more commonly used as 1 Pa is a very small pressure.

What is the conversion factor between kPa and mmHg?

1 kPa = 7.501 mmHg (equivalently, 1 mmHg ≈ 0.133 kPa). This conversion is clinically relevant when interpreting arterial blood gas partial pressures.

Classify the methods of measuring pressure used in anaesthesia.
  • Direct methods: liquid-containing gauges (manometer, barometer), aneroid gauges (Bourdon gauge, bellows), electronic transduction (strain gauge)
  • Indirect methods: non-invasive blood pressure cuff (oscillometric), Penaz technique (continuous finger plethysmography), Doppler ultrasound
What is pulse oximetry and what parameter does it directly measure?

Pulse oximetry is a non-invasive optical technique that measures oxygen saturation of haemoglobin in pulsatile arterial blood (SpO₂). It uses two wavelengths of light (660 nm red and 940 nm infrared) transmitted through a tissue bed; the ratio of absorbances at these wavelengths is compared against an empirical calibration curve derived from healthy volunteers.

Start free trial