Home  /  RANZCR Part 1  /  Study notes  /  Quality Assurance and Quality Control in Medical Imaging

Quality Assurance and Quality Control in Medical Imaging

RANZCR Part 1 LO 2.3.41 2,680 words
Free preview. This study note maps to learning objective 2.3.41 in the RANZCR Part 1 curriculum. Inside Primex you get the full set of RANZCR Part 1 notes, AI-graded SAQs and written-paper practice, voice viva with an AI examiner, exam-style MCQs, and a curriculum tracker that ticks off every learning objective as you go. For exam format, timeline and failure-mode commentary, see the RANZCR Part 1 2026 Study Guide.

Overview and Principles of Quality Assurance

Quality assurance (QA) in medical imaging encompasses the entire system of management policies, procedures, and processes implemented to ensure that every imaging examination is appropriate, technically optimal, correctly interpreted, communicated in a timely manner, and delivered at the lowest practicable radiation dose. QA is broader than quality control (QC): it includes clinical governance, credentialing, continuing medical education, protocol optimisation, equipment maintenance, preventive maintenance, calibration, and performance monitoring across the full imaging chain.

Quality control is the subset of QA concerned specifically with the technical performance of equipment. QC involves systematic, periodic testing of imaging systems against defined standards to identify drift, degradation, or failure before these translate into suboptimal clinical images or patient harm.

Core Principles of QA

Principle Description
Appropriateness Ensuring each examination is clinically justified
Optimisation (ALARA) Minimising dose while maintaining diagnostic adequacy
Accuracy Verifying that measurements, doses, and images meet specifications
Consistency Reproducible performance over time and across operators
Documentation Formal recording of tests, results, and corrective actions
Accountability Defined responsibility (lead radiologist/physicist/RSO)

Benefits of QA


Increased QA Requirements for Asymptomatic (Screening) Populations

Screening programmes differ fundamentally from diagnostic imaging in that they are applied to large populations of healthy, asymptomatic individuals where the pre-test probability of disease is low. This has critical implications for the required QA framework.

Why Stricter QA Is Needed in Screening

  1. Narrower harm-to-benefit ratio: Asymptomatic individuals receive no immediate therapeutic benefit; any avoidable dose, false positive, or missed lesion represents a net harm.
  2. High volume, low prevalence: Even small degradations in sensitivity or specificity have large population-level consequences - increased false negatives yield missed cancers; false positives yield unnecessary biopsies and anxiety.
  3. Minimum detectable lesion size is critical: Screening depends on detecting early-stage, often small lesions (e.g., microcalcification clusters, architectural distortion in mammography); equipment not performing to specification will miss these.
  4. Regulatory requirements are heightened: Mammography screening is one of the most stringently regulated imaging applications. In the United States, the Mammography Quality Standards Act (MQSA, enacted 1992, CFR Title 21 Part 900) makes many ACR QA procedures mandatory, making mammography the only federally regulated imaging modality in the US. Accreditation and certification by the FDA are required before a facility may provide mammography services.
  5. Programme-level auditing: Recall rates, cancer detection rates, positive predictive values, and interval cancer rates must be tracked as QA outcome measures beyond equipment testing.
  6. Clinical image quality feedback is mandated: In mammography, the FDA's EQUIP initiative (launched January 2017) requires the Lead Interpreting Radiologist to provide ongoing feedback on image quality, document corrective actions, and conduct regular technologist image quality reviews.

Screening-Specific QA Measures

Screening Modality Key QA Requirement
Mammography Category A/B/C QC tests; breast dose limits; accreditation phantom imaging; SNR/CNR thresholds; DRL monitoring
Lung CT (low-dose) CTDI monitoring; nodule volumetry accuracy; spatial resolution verification
CT colonography Noise index protocols; lumen distension phantom testing
Ultrasound adjunct (breast) Transducer performance; penetration depth; spatial resolution testing

For mammography, QC tests are categorised under the regulatory framework:

For digital mammography and digital breast tomosynthesis (DBT), manufacturers' QC procedures must be followed; an FDA-approved manufacturer-independent harmonised QC programme is available through the 2018 ACR Digital Mammography Quality Control Manual.


Quality Control Testing by Imaging Modality

Radiographic and Fluoroscopic Equipment QC

Radiographic QC is performed at acceptance testing, after any major repair, and at scheduled intervals (typically annually by a medical physicist, with more frequent technologist-level checks).

Test Parameter Assessed Frequency
kVp accuracy Generator output voltage (typically ±5%) Acceptance, annually
Timer accuracy Exposure time accuracy Acceptance, annually
mA linearity Proportionality of output to mA setting Acceptance, annually
Radiation output reproducibility Consistency of output at fixed settings Acceptance, annually
Half-value layer (HVL) Beam quality / adequacy of filtration Acceptance, annually
Spatial resolution Line-pair or star phantoms; limiting spatial frequency Acceptance, annually
Contrast resolution Low-contrast phantom objects Acceptance, annually
Collimation / light field alignment Correspondence of light field and radiation field Acceptance, annually
AEC (automatic exposure control) Consistency of detector dose across thickness/kVp Acceptance, annually
Flat-field uniformity Detector uniformity Acceptance, annually
Reject / retake analysis Percentage of repeat examinations and cause Ongoing
DRL monitoring Entrance air kerma / dose benchmarking Ongoing

The half-value layer is defined as:

$$\text{HVL} = \frac{\ln 2}{\mu}$$

where $\mu$ is the linear attenuation coefficient of the attenuating material (typically aluminium). HVL measurement verifies beam hardening and filtration adequacy; inadequate filtration increases patient dose without diagnostic benefit.

Spatial resolution testing uses line-pair phantoms or star patterns. The observer identifies the limiting spatial frequency (reported in $\text{lp} \cdot \text{mm}^{-1}$) as the smallest resolvable bar group. For fluoroscopic systems, additional tests include dose rate, frame rate, and automatic brightness control performance.

Reject/retake analysis is an important ongoing QC component: portable chests, lumbar and thoracic spines, KUBs, and abdomens typically have the highest retake rates. Digital receptors (photostimulable phosphor plates or direct digital detectors) substantially reduce retake rates compared with film-screen systems.


Nuclear Medicine Equipment QC

Gamma Camera (Planar and SPECT)

QC of the Anger scintillation camera is essential because non-uniformity, non-linearity, and energy resolution errors introduce systematic artefacts into planar and tomographic images. In SPECT, even small (1-2%) non-uniformities are amplified into ring artefacts by the reconstruction process; flood correction maps must therefore be updated regularly.

Test Parameter Assessed Frequency
Daily flood (uniformity) Intrinsic/extrinsic field uniformity Daily
Energy peaking Photopeak centring for radionuclide in use Daily (with source change)
Spatial resolution (bar phantom) System resolution Weekly
Spatial linearity Geometric distortion Weekly
Centre of rotation (SPECT) Rotational axis alignment Monthly or after repair
Sensitivity Counts per unit activity Quarterly
Multiple window spatial registration Window alignment (multi-isotope studies) Quarterly

Dose Calibrator QC

The dose calibrator assay is often the only check that a patient receives the prescribed radiopharmaceutical activity; its QC is therefore mandated by the NRC and state agencies, in accordance with nationally recognised standards or the manufacturer's instructions.

Test Standard Frequency
Accuracy Measured activity within 10% of certified reference sources (e.g., $^{57}$Co, $^{137}$Cs, activities known to within 5%) Acceptance, annually
Linearity Activity accuracy across the clinical range of activities Acceptance, quarterly
Geometry Constancy for different volumes/container types Acceptance; if geometry changes
Constancy Day-to-day reproducibility Daily

Note: attenuation correction factors for specific radionuclides, containers, and dose calibrator models may be required for accurate assay; this is an unresolved challenge for some radionuclides.

PET and PET/CT QC

Acceptance testing follows the NEMA NU2-2018 standard, which defines methodology for: 1. Spatial resolution 2. Scatter fraction, count losses, and randoms 3. Sensitivity 4. Accuracy of corrections for count losses and randoms 5. Image quality, accuracy, and corrections 6. Time-of-flight (ToF) resolution (where applicable) 7. PET/CT co-registration accuracy (hybrid systems)

Results are compared against manufacturer specifications.

Test Frequency
Daily normalisation / blank scan Daily
Energy and timing resolution Daily
Detector efficiency normalisation After detector module replacement or gantry opening
Spatial resolution Acceptance, annually
Sensitivity Acceptance, annually
Scatter fraction and count-rate performance Acceptance, annually
Image quality (ACR PET phantom) Per accreditation schedule
PET/CT co-registration accuracy After hardware change; periodically
CT component QC Per standard CT QC schedule

Routine testing schedules are additionally specified by AAPM Report No. 126 (PET/CT Acceptance Testing and QA, 2019) and individual manufacturer recommendations.

ACR PET accreditation phantom - three sections:

Section Contents Purpose
Top Four "hot" cylinders (8, 16, 22, 25 mm diameter); three 25-mm cylinders of air, water, and Teflon Image contrast (max SUV); scatter/attenuation correction verification (mean/min SUV)
Middle Uniform activity fill Image uniformity
Bottom Six groups of variable-diameter plastic cylinders in pie arrangement Spatial resolution

All measurements are made on images reconstructed with full corrections applied (attenuation, scatter, randoms, dead time).


MRI Equipment QC

MRI QC must cover the entire imaging chain: the main magnetic field ($B_0$), radiofrequency (RF) system, gradient system, receiver coils, and display systems. The ACR MRI accreditation programme specifies minimum QC requirements, assessing personnel qualifications, equipment performance, QC procedure effectiveness, and clinical image quality.

Key components requiring periodic monitoring: - Magnetic field strength and homogeneity (shimming effectiveness) - Gradient linearity and calibration (affects geometric accuracy and slice profile) - RF tuning and transmitter gain (affects flip angle accuracy and SNR) - Receiver coil impedance matching and performance - Environmental noise sources and power supplies

Test Parameter Frequency
Central frequency Main field stability ($B_0$) Daily/weekly
Signal-to-noise ratio (SNR) Receiver chain performance Weekly
Image uniformity RF field ($B_1$) homogeneity Weekly
Geometric accuracy Gradient linearity, dimensional accuracy Weekly
Spatial resolution High-contrast resolution phantom Weekly
Low-contrast object detectability Contrast resolution Weekly
Ghosting / artefact assessment Gradient and RF artefacts Weekly
Slice position accuracy Gradient calibration Weekly
Slice thickness accuracy Profile width Weekly
RF coil performance SNR per clinical coil Acceptance, annually

ACR MRI accreditation phantom: cylindrical vessel, 190 mm internal diameter × 148 mm internal length, filled with a nickel sulphate/sodium chloride aqueous solution adjusted to provide clinically relevant T1 and T2 relaxation times. Contains geometric arrays for spatial accuracy, resolution inserts, and structures for slice position and thickness verification.

MRI phantoms more generally use aqueous paramagnetic solutions, gelatin, agar, silicone, agarose, or organically/paramagnetically doped gels. Dopants (nickel, manganese, gadolinium, aqueous oxygen) are used to adjust T1 and T2 to values appropriate for clinical pulse sequence timing.

A comprehensive systems test - assessing spatial resolution, statistical noise, count-rate performance, sensitivity, and image quality - should be performed at least annually.


Ultrasound Equipment QC

Ultrasound QC relies on tissue-mimicking phantoms and structured performance testing, complemented by daily informal assessment by the sonographer during routine clinical use. QC is essentially performed every day during routine scanning by the sonographer, who should recognise major problems; however, a formal periodic QC programme is required to detect problems before serious malfunctions occur.

Tissue-mimicking phantom material: attenuation coefficient of approximately $0.5$ to $0.7~\text{dB} \cdot \text{cm}^{-1} \cdot \text{MHz}^{-1}$ (the higher value provides a more challenging penetration test) and a speed of sound representative of soft tissue.

A standard general-purpose phantom comprises three modules:

Module Contents Tests Performed
Resolution/geometry High-contrast pin targets at calibrated depths; vertical and horizontal calibration arrays Axial and lateral spatial resolution; horizontal and vertical distance accuracy; dead zone depth
Elevational resolution Small-diameter sphere/cylinder targets (2 and 4 mm) at varying depths Slice-thickness (elevational) resolution vs. depth
Uniformity/penetration Uniformly distributed scattering material Image uniformity; maximum penetration depth

Additional low-contrast sphere inserts assess contrast resolution.

QC Test What is Measured
Axial spatial resolution Along-beam resolution
Lateral spatial resolution In-plane, perpendicular-to-beam resolution
Elevational (slice-thickness) resolution Out-of-plane resolution
Depth of penetration Maximum imaging depth
Image uniformity Gain homogeneity; dead/malfunctioning elements
Geometric accuracy Horizontal and vertical distance accuracy
Dead zone Non-imaged region immediately adjacent to transducer face
Contrast resolution Detection of low-contrast structures
Doppler performance Velocity accuracy; PRF calibration

Each transducer and transducer port must be tested individually. Additional system capabilities (harmonic imaging, elastography, colour flow Doppler) must be evaluated and documented to meet published specifications at acceptance. Transducer surface integrity should be checked for delamination or cracked face.

Testing frequency: establish empirically - perform tests frequently initially, review logbooks over an extended period, and with documented stability, reduce the testing rate as appropriate. This risk-informed approach allocates QC effort where instability is most likely.

Power output (acoustic output) of transducers can be assessed by weighing sound pressure with a force balance or measuring heating effect using a calorimeter; more sophisticated methods characterise the intensity or pressure distribution across the field.


Roles and Responsibilities in QA

Role Responsibilities
Lead Radiologist Overall accountability for the QA programme; clinical image quality review; corrective action oversight; mandatory in mammography (EQUIP)
Diagnostic Medical Physicist Acceptance testing; annual surveys; QC programme design; protocol optimisation; shielding calculations; post-repair testing
Radiation Safety Officer (RSO) Regulatory compliance; staff dose monitoring; licence management; patient safety regarding radiation and dosing
Radiographer / Technologist Daily QC tests; reject analysis; artefact identification; immediate fault reporting
Service Engineer Hardware repair; post-repair testing in conjunction with physicist

The diagnostic medical physicist is responsible for testing imaging equipment initially, annually, and after repairs or modifications that may affect radiation doses to patients or image quality. This includes dose calibrator testing, gamma camera performance surveys, MRI system surveys, and CT dose verification.


Acceptance Testing vs. Routine QC vs. Annual Survey

Phase Timing Purpose
Acceptance testing On installation or relocation Verify manufacturer specifications are met; establish performance baseline
Routine QC Daily to quarterly, depending on test Detect drift or degradation early
Annual survey (medical physicist) Yearly; or after major repair Comprehensive performance review; DRL comparison; protocol audit

Acceptance testing is the most comprehensive phase. For all modalities it establishes the baseline against which subsequent QC measurements are compared. Parameters outside manufacturer-specified tolerance at acceptance must be resolved before the equipment enters clinical use.


Documentation and Corrective Action

An effective QA programme requires:

  1. Written programme document: goals, responsibilities, test procedures, equipment, tolerances, and corrective action thresholds
  2. Logbooks: dated records of all QC measurements, the operator, and equipment status
  3. Corrective action log: problem identified, investigation performed, action taken, verification of resolution, and date of return to service
  4. Trend analysis: graphical display of QC metrics over time to detect gradual drift before action limits are breached
  5. Clinical image quality feedback loop: radiologists must have a formal mechanism to report image quality concerns; in screening this is especially important as missed lesions can drive equipment re-evaluation

For mammography, the EQUIP framework mandates that the Lead Interpreting Radiologist review all required performance tests, document corrective actions, engage in regular image quality audits of technologist performance, and ensure mechanisms are in place for ongoing radiologist feedback on image quality - embedding clinical feedback within the technical QC cycle.


Summary: Key QC Parameters by Modality

Modality Critical QC Parameters
Radiography / fluoroscopy kVp, HVL, mA linearity, AEC, spatial resolution, flat-field uniformity, reject/retake rate, DRL
Mammography (FFDM / DBT) Breast dose, phantom image quality (ACR), SNR/CNR, spatial resolution, AEC, compression paddle, flat-field calibration (Category A/B/C framework)
CT CTDI$_\text{vol}$, noise index, spatial/contrast resolution, HU accuracy, table position accuracy, DRL
Gamma camera / SPECT Flood uniformity, energy resolution, spatial resolution and linearity, centre of rotation
Dose calibrator Accuracy, linearity, geometry, constancy
PET / PET-CT Blank scan, normalisation, spatial resolution, sensitivity, scatter fraction, image quality (ACR phantom), PET-CT co-registration; NEMA NU2-2018 acceptance standard
MRI SNR, image uniformity, geometric accuracy, spatial resolution, ghosting, slice position and thickness accuracy, coil performance
Ultrasound Axial/lateral/elevational spatial resolution, penetration depth, image uniformity, geometric accuracy, dead zone, contrast resolution, Doppler velocity accuracy

A robust QA programme across all imaging modalities ensures that patients - both symptomatic and asymptomatic - receive examinations that are diagnostically adequate, dose-optimised, and performed on equipment proven to meet defined performance standards. For screening populations in particular, the stakes of equipment underperformance are amplified across large numbers of healthy individuals, making rigorous, well-documented, and regularly reviewed QC essential to programme integrity.

Primex

Practice this topic in the app

Work through MCQs on this exact LO, run written or viva practice mapped to 2.3.41, or ask PRIMEX a clinical question framed for RANZCR Part 1. Your free trial covers all 20 specialist exams.

Start 7-day free trial
Start free trial